8.7 Command Message Description
233
8
Chapter 8
Contro
lling the
Instru
ment From
a Comp
uter
Function
Query
Returns the content of the panel of the specified number.
Example
Query
Response
:SYSTem:FILE? 1
:SYSTEM:FILE ABC,NO-111,CLA1,BF,B2,12,0,2,0,1,
+100.0E-06,+500.0E-06,+10.00E-06,+50.00E-06,
0,0,1,0,1,1,2,0,1,0,0,0,0,0,0,0
(Headers: ON)
ABC,NO-111,CLA1,BF,B2,12,0,2,0,1,+100.0E-06,
+500.0E-06,+10.00E-06,+50.00E-06,0,0,1,0,1,1,2,
0,1,0,0,0,0,0,0,0
(Headers: OFF)
The following settings have been made:
<Equipment name>
: ABC
<Control number>
: NO-111
<Grounding class>
: Class-I equipment
<Applied part>
: BF-type applied
parts
<Network>
: Network B2
<Measurement mode>
: Patient leakage
current:SIP/SOP
<Measurement method>
: Manual
<Target current>
: DC
<Measurement range>
: Auto range
<Filter>
: ON
<Allowable value in normal condition
(upper limit)>
: 0.1 mA
<Allowable value in single-fault condition
(upper limit)>
: 0.5 mA
<Normal condition allowable value
(upper limit, DC measurements)>
: 0.01mA
<Single fault-condition allowable value
(upper limit, DC measurements
)
> :
0.05mA
<Allowable value in normal condition
(lower limit ON/OFF)>
: OFF
<Allowable value in single-fault condition
(lower limit ON/OFF)>
: OFF
<Normal condition allowable value
(lower limit ON/OFF, AC measurements)>
: ON
<Single fault-condition allowable value
(lower limit ON/OFF, AC measurements)>
: OFF
<Equipment under test status>
: Single-fault
condition
Disconnection of one wire in power line)
<Power supply polarity>
: Negative phase
<Other 110% voltage application>
: 110% voltage
application:
Negative phase
<Specific 110% voltage application>
: 110% voltage
application:None
<Switch>
: S10=ON,S12=OFF,S13=OFF
<Measurement item in automatic measurement>
: None
<Measurement item in automatic measurement2>
: None
<Measuring time in automatic measurement>
: None
<Wait time in automatic measurement(delay a)>
: None
<Wait time in automatic measurement(delay b2)>
: None
<Wait time in automatic measurement(delay c)>
: None
<Wait time in automatic measurement(delay b1)>
: None
Error
If the designated panel number does not conform to the specified range, an
execution error occurs.
Note
Returns 0 in the following cases:
•
<numerical value 11>
(Equipment under test status)
<numerical value
12>
(Power supply polarity) in automatic measurement
<numerical value 13>
(Other 110% voltage application)
<numerical value
14>
(
Specific 110% voltage application
)
<numerical value 15>
(Switch)
•
<numerical value 16>
to
<numerical value 22>
in manual measurement
• When there is no data
Queries the content of a panel
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......