3.3 Connecting equipment under test to the instrument
37
3
Chapte
r 3 Setting Prep
arations
Using a wiring adapter
Obtain and use a wiring adapter as shown below.
Connecting wires to the terminal block
GN L
Check that the G, N and L terminals
are correctly connected and make
sure the wires are properly secured
Turn on the instrument.
Fit the right and left projecting por-
tions of the terminal block cover
into the panel slit. Slide the cover
downwards until it clicks into place
Slide upwards
Raise the cover.
(The cover will stay in
place when raised to a
certain height)
G : For grounding wire connection
(Class-I equipment only)
N : For neutral wire connection
L : For live wire connection
Turn off the instrument and set the breaker to Off.
See
See
"3.1 Power switch, breaker ON/OFF" (p. 31)
1
2
3
4
5
Turn on the HiTESTER.
See
"3.4 Turning Power On and OFF" (p. 38)
6
Clicking
sound
• Turn off the circuit breaker connecting a wiring adapter or equipment.
• The terminal block continuously output the voltage based on the power sup-
plied to [LINE IN].
Multi-connection adapter, etc.
Leakage current will increase with the use of multi-connection adapters, etc.
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......