Appendix3 List of default settings
A21
Appendix
Total patient
leakage
current
( F-type ap-
plied part)
measurement
–
manual
–
manual
–
manual
polarity positive
phase
positive
phase
–
equipment
status
–
–
–
Other applied
voltages
–
–
–
Application of
specific voltage
110%
applied N
110%
applied N
110%
applied N
s10
OFF
OFF
–
s12
–
–
–
s13
OFF
OFF
–
filter
ON
ON
ON
target current
ACDC
ACDC
ACDC
range
AUTO
AUTO
AUTO
allowable value
: upper limit:
normal
–
–
–
allowable value:
upper limit: fault
5.000
mA
100.0
A
5.000
mA
100.0
A
5.000
mA
100.0
A
allowable value:
lower limit :
ON/OFF:
normal
–
–
–
allowable value:
lower limit: ON/
OFF: fault
OFF
OFF
OFF
Total patient
leakage
current
(metal ac-
cessible
part)
measurement
manual
–
manual
–
manual
–
polarity positive
phase
positive
phase
–
equipment
status
–
–
–
Other applied
voltages
–
–
–
Application of
specific voltage
110%
applied N
110%
applied N
110%
applied N
s10
OFF
OFF
–
s12
–
–
–
s13
–
–
–
filter
ON
ON
ON
target current
ACDC
ACDC
ACDC
range
AUTO
AUTO
AUTO
allowable value
: upper limit :
normal
–
–
–
allowable value
: upper limit :
fault
1.000 mA
1.000 mA
1.000 mA
allowable value
: lower limit :
ON/OFF :
normal
–
–
–
allowable value
: lower limit :
ON/OFF : fault
OFF
OFF
OFF
default setting
Class I
Class II
Internally powered
Leakage
current mode
item
Type B
Type BF Type CF Type B
Type BF Type CF Type B
Type BF Type CF
*
( ) is the value after DC switching
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......