
10.2 Auxiliary measurement functions*
256
10.2 Auxiliary measurement functions*
110% voltage
application
function
Equipped with a voltage output terminal (T3) that applies a voltage that is 110% of rated
voltage between the insulated SIP/SOP section, F-type applied part, or patient
connections and ground.
No-load output voltage:
From -0% to +5% of power supply voltage of line input for equipment under test
1 M
resistance load:
Within ± 2% of power supply voltage of line input for equipment under test
Output impedance: 22.5 ± 3 k
(including output protective resistance of 10 k
at 50/60 Hz)
Output ON/OFF selection
(a)Applied when positive phase (to input power supply voltage)
(b)Applied when negative phase (to input power supply voltage)
It is possible to automatically switch between (a) and (b) (automatic testing function)
*
only ST5540
10.3 Measurement System
Measurement
system
Indication of a current value calculated based the measured drop in voltage
caused by simulated resistance of the human body
Measurement of true effective value
The measurement section : chassis-grounded and floating
Simulated resistance
of human body
(current detection circuit)
Resistance tolerance of
used elements:
Resistance:±1%
Capacitor: ±3%
Network A
For Electrical
Appliance and
Material Safety
Law
Basic measurement element :1 k
Filter: 10 k
+ 11.22 nF + 579
Network B1, B2*
For medical
electrical
equipment
Basic measurement element :1 k
Filter: 10 k
+ 15 nF
Network C
For IEC 60990
Basic measurement element :
1.5 k
+ 500
Filter1:10 k
+ 22 nF
Filter2:
10 k
+ (20 k
+ 6.2 nF)//9.1 nF
Network D
For UL
Basic measurement element :
1.5 k
//0.15
F
Network E
General-purpose 1 Basic measurement element1 :1 k
Network F
General-purpose 2 Basic measurement element2 :2 k
Network G
For IEC 60601-1
For damp
conditions
Basic measurement element :
375
//0.22
F + 500
Protective
conductor
current
Basic measurement element :35
Measurement procedure
Manual
measurement
Measure each item while changing the measurement
conditions
Automatic
measurement
Automatic measurement of power supply polarity and
equipment status
Measurement time setting
Set the delay (wait time) for state transitions
Measurement terminals
Terminal T1, terminal T2 (with built-in fuse holder),
Terminal T3 (110% voltage application terminal)*
*
only ST5540
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......