9.2 Connecting to the EXT I/O Terminal
248
Table of LOAD0 - LOAD4 control and corresponding panel numbers
INT.DCV
INT.GND
Output
Outputs the instrument's internal 5 VDC and GND.
EXT.DCV
EXT.COM
Input
Terminal for power input from external equipment
Allowed input voltage range: +5 to +24 VDC
(Reserved)
-
Do not use this terminal.
Terminal
name
Input/
output
Function
Number of panel
to be loaded
LOAD4
LOAD3
LOAD2
LOAD1
LOAD0
1
1
1
1
1
0
2
1
1
1
0
1
3
1
1
1
0
0
4
1
1
0
1
1
5
1
1
0
1
0
6
1
1
0
0
1
7
1
1
0
0
0
8
1
0
1
1
1
9
1
0
1
1
0
10
1
0
1
0
1
11
1
0
1
0
0
12
1
0
0
1
1
13
1
0
0
1
0
14
1
0
0
0
1
15
1
0
0
0
0
16
0
1
1
1
1
17
0
1
1
1
0
18
0
1
1
0
1
19
0
1
1
0
0
20
0
1
0
1
1
21
0
1
0
1
0
22
0
1
0
0
1
23
0
1
0
0
0
24
0
0
1
1
1
25
0
0
1
1
0
26
0
0
1
0
1
27
0
0
1
0
0
28
0
0
0
1
1
29
0
0
0
1
0
30
0
0
0
0
1
1:High level, 2:Low level
Invalid setting results when all terminals from LOAD0 to LOAD4 are set to "1" or
"0."
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......