8.7 Command Message Description
206
Example
Command
:CONFigure:AUTO ON;:CONFigure:WTIMe:POLarity
0
10
Sets the wait time (delay b1) for switching operation in automatic measurement
to 10 seconds.
Query
Response
:CONFigure:WTIMe:POLarity
0
?
:CONFIGURE:WTIME:POLARITY
0
1200
(Headers : ON)
1200
(Headers : OFF)
The wait time (delay b1) for switching operation in automatic measurement has
been set to 1,200 seconds.
Error
An execution error occurs in the following cases :
• When the measurement method is set to "manual measurement"
• When the measurement mode is not selected
• When a mode other than "leakage current mode" is set
• In automatic measurement (Command)
• When a value outside the specified range is input
Note
• Use this command after executing the
:MODE
command to activate the mea-
surement mode.
• The unit is [s]. Fractions are rounded up when 0.5 or greater and rounded off
when less than 0.5.
Sets and queries the wait time for switching operation in automatic measure-
ment (delay b1)
Sets and queries the grounding class of equipment under test
Syntax
Command
Query
Response
:EQUipment
<characters>
:EQUipment?
<characters>
<characters>
= CLAss1(CLA1)/ CLAss2(CLA2)/ INTernal
CLAss1 : Class-I equipment
CLAss2 : Class-II equipment
INTernal: Internally powered equipment
Function
Command
Sets the grounding class of equipment under test.
Query
Returns the grounding class setting of equipment under test as alphabetic char-
acters.
Example
Command
:MODE OFF;:EQUipment CLAss1
Sets the grounding class of equipment under test to "Class-I equipment."
Query
Response
:EQUipment?
:EQUIPMENT CLASS1
(Headers : ON)
CLASS1
(Headers : OFF)
The grounding class of equipment under test has been set to "Class-I equip-
ment."
Error
An execution error occurs in the following cases:
• When the measurement mode is selected (Command)
• When a mode other than "leakage current mode" is set (Command)
Summary of Contents for ST5540
Page 2: ......
Page 6: ...Contents iv Index i ...
Page 29: ...23 2 Chapter 2 Instrument Labels and Screens Instrument Labels and Screens Chapter 2 ...
Page 80: ...4 8 Changing the Measurement Method Auto Manual 74 ...
Page 121: ...115 8 9 7 Chapter 7 Instrument System Settings Instrument System Settings Chapter 7 ...
Page 144: ...7 14 Setting the Backlight 138 ...
Page 284: ...10 14 Measurement Networks 278 ...
Page 290: ...11 4 Instrument Disposal 284 ...
Page 320: ...Appendix4 External Dimensions A30 ...
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......