10
Field of view.
%
Humidity limitation: relative humidity between
30 and 85 % in operating environment.
Atmospheric pressure limitation: between 80 and
109 kPa in operating environment.
Electrical Safety Type BF Applied Part.
Packaging level ensuring sterility.
UL Recognized Component Mark for Canada and
the United States.
Medical device.
Global Trade Item Number.
Country of manufacturer.
INMETRO Certificate Medical Electrical Equipment
ABNT NBR IEC 60601-1, ABNT NBR IEC 60601-1-2,
ABNT NBR IEC 60601-2-18.
Applicable for Ambu® aScope™ 4 Broncho and
Ambu® aView™ monitor.
Do not use if the product sterilisation barrier
or its packaging is damaged.
4. Use of aScope 4 Broncho Sampler Set
Numbers in gray circles below refer to illustrations on page 2.
4.1. Preparation and inspection of aScope 4 Broncho Sampler Set
Visual inspection of the endoscope 1
1. Check that the pouch seal is intact.
1a
2. Make sure to remove the protective elements from the handle and from the
insertion cord.
1b
3. Check that there are no impurities or damage on the product such as rough surfaces,
sharp edges or protrusions which may harm the patient.
1c
Refer to the Ambu displaying unit
Instruction for use
for preparation and inspection
of the Ambu displaying unit 2
Inspection of the image
1. Plug in the endoscope cable connector into the corresponding connector on the
compatible displaying unit. Please ensure the colours are identical and be careful to align
the arrows.
3
2. Verify that a live video image appears on the screen by pointing the distal tip of the
endoscope towards an object, e.g. the palm of your hand.
4
3. Adjust the image preferences on the displaying unit if necessary (please refer to the
displaying unit
Instruction for use
).
4. If the object cannot be seen clearly, wipe the lens at the distal tip using a sterile cloth.
Preparation of aScope 4 Broncho Sampler Set
1. Carefully slide the control lever forwards and backwards to bend the bending section as much
as possible. Then slide the control lever slowly to its neutral position. Confirm that the bending
section functions smoothly and correctly and returns to a neutral position.
5a
Summary of Contents for aScope 4 Broncho Large
Page 17: ...BG 17 5 4 6 6 4 1 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Ambu Ambu 24...
Page 18: ...18 25 26 27 28 1 2 3 4 ETT 5 6 7 8 9 10 85 kPa 638 mmHg 11 1 5...
Page 21: ...BG 21 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Luer Lock B 15 16 17 18...
Page 28: ...28 6 2 2...
Page 29: ...BG 29 aScope 4 Broncho aScope 4 Broncho...
Page 69: ...EL 69 5 4 6 6 4 1 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Ambu Ambu 24...
Page 70: ...70 25 26 27 28 1 2 3 4 5 6 7 8 9 10 85 kPa 638 mmHg 11 1 5 1 6...
Page 73: ...EL 73 A aScope 4 Broncho 1 2 3 4 5 6 7 8 9 LED 10 11 12 13 14 Luer Lock B 15 flip top 16...
Page 80: ...80 6 2 2...
Page 81: ...EL 81 aScope 4 Broncho aScope 4 Broncho...
Page 180: ...180 aScope 4 Broncho aScope 4 Broncho...
Page 269: ...RU 269 5 4 6 6 4 1 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Ambu Ambu...
Page 270: ...270 24 25 26 27 28 1 2 3 4 ETT 5 6 7 8 9 10 85 638 11 1 5...
Page 273: ...RU 273 1 2 3 4 5 6 7 8 9 10 11 12 13 14 B 15 16 17 18 C...
Page 280: ...280 6 2 2...
Page 281: ...RU 281 aScope 4 Broncho aScope 4 Broncho...
Page 339: ...ZH 339 6 2 2 aScope 4 Broncho aScope 4 Broncho...