12
Insertion of endoscopic accessories 7d
Always make sure to select the correct size endoscopic accessory for the endoscope.
Inspect the endoscopic accessory before using it. If there is any irregularity in its operation or
external appearance, replace it. Insert the endoscopic accessory into the working channel port
and advance it carefully through the working channel until it can be seen on the displaying
unit. The enclosed introducer can be used to facilitate the insertion of soft accessories such as
microbiology brushes.
Operating with the Broncho Sampler Set
1. If suction is deemed relevant before sampling, ensure that the flow switch is aligned
vertically and pointing upwards, and perform suction by pressing the suction button on
the aScope 4 Broncho in accordance with
Instruction for use
for aScope 4 Broncho.
7e
2. To prepare for sampling of specimen, turn the flow switch counter-clockwise towards the
sample container, and align it horizontally.
7f
3. Perform sampling by pressing the suction button on the aScope 4 Broncho. Do not fill the
sample container above the maximum of the volume indicator scale.
4. If suction is deemed relevant, turn the flow switch clockwise to align it vertically and
pointing upwards, and press the suction button on the aScope 4 Broncho. When suction
is complete, turn the flow regulator counter-clockwise towards the sample container, and
align it horizontally.
5. Remove the sample container by pressing the flip top, and pull the sample container from
the sampler bridge.
7g
6. Close the sample container by pressing the sample container lid towards the sample
container interface and engage the click-lock. Once the click-lock is engaged, do not
re-open the lid.
7h
Withdrawal of the aScope 4 Broncho Sampler Set
1. If deemed relevant, detach the sampler bridge from the aScope 4 Broncho by disengaging
the attachment lock, and pull the sampler bridge off the aScope 4 Broncho. Disengage
suction by pulling the suction tube off the sampler bridge.
2. If deemed relevant, the sample can be accessed by unscrewing the sample container glass
counter-clockwise. It is recommended to handle contaminated substances in accordance
with local guidelines.
8a
3. Transportation and handling of sample container must be performed in accordance with
local guidelines for transportation and handling of hazardous material.
4. When withdrawing the aScope 4 Broncho, make sure that the control lever is in the
neutral position. Slowly withdraw the endoscope while watching the live image on the
displaying unit.
8b
4.3. After use
Visual check 9
1. Are there any missing parts on the bending section, lens, or insertion cord? If yes, then
take corrective action to locate the missing part(s).
2. Is there any evidence of damage on the bending section, lens, or insertion cord? If yes,
then examine the integrity of the product and conclude if there are any missing parts.
3. Are there cuts, holes, sagging, swelling or other irregularities on the bending section, lens,
or insertion cord? If yes, then examine the product to conclude if there are any missing parts.
In case of corrective actions needed (step 1 to 3) act according to local hospital procedures.
The elements of the insertion cord are radio opaque.
Disconnect
Disconnect the endoscope from the displaying unit
10
. The aScope 4 Broncho Sampler Set is a
single use device. The aScope 4 Broncho Sampler Set is considered infected after use and must
be disposed of in accordance with local guidelines for collection of infected medical devices
with electronic components. Do not soak rinse, or sterilize this device as these procedures may
leave harmful residues or cause malfunction of the device. The design and material used are
not compatible with conventional cleaning and sterilization procedures.
Summary of Contents for aScope 4 Broncho Large
Page 17: ...BG 17 5 4 6 6 4 1 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Ambu Ambu 24...
Page 18: ...18 25 26 27 28 1 2 3 4 ETT 5 6 7 8 9 10 85 kPa 638 mmHg 11 1 5...
Page 21: ...BG 21 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Luer Lock B 15 16 17 18...
Page 28: ...28 6 2 2...
Page 29: ...BG 29 aScope 4 Broncho aScope 4 Broncho...
Page 69: ...EL 69 5 4 6 6 4 1 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Ambu Ambu 24...
Page 70: ...70 25 26 27 28 1 2 3 4 5 6 7 8 9 10 85 kPa 638 mmHg 11 1 5 1 6...
Page 73: ...EL 73 A aScope 4 Broncho 1 2 3 4 5 6 7 8 9 LED 10 11 12 13 14 Luer Lock B 15 flip top 16...
Page 80: ...80 6 2 2...
Page 81: ...EL 81 aScope 4 Broncho aScope 4 Broncho...
Page 180: ...180 aScope 4 Broncho aScope 4 Broncho...
Page 269: ...RU 269 5 4 6 6 4 1 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Ambu Ambu...
Page 270: ...270 24 25 26 27 28 1 2 3 4 ETT 5 6 7 8 9 10 85 638 11 1 5...
Page 273: ...RU 273 1 2 3 4 5 6 7 8 9 10 11 12 13 14 B 15 16 17 18 C...
Page 280: ...280 6 2 2...
Page 281: ...RU 281 aScope 4 Broncho aScope 4 Broncho...
Page 339: ...ZH 339 6 2 2 aScope 4 Broncho aScope 4 Broncho...