14
Suction connector
Connecting tube ID [mm]
Ø6.5 - 9.5
Connecting tube with female interface
ID = Ø7 ± 1 mm
Connecting tube with male interface.
Applicable when suction adapter is used
OD = Ø7 ± 1 mm
1. Please check for compatibility with suction tubing before use
Sample container
Specification
Sample Container volume
30 ml
Sterilisation
aScope 4 Broncho Regular/Large Sampler Set
Method of sterilisation
ETO
Operating environment
aScope 4 Broncho Regular/Large Sampler Set
Temperature [°C, (°F)]
10 ~ 40 (50 ~ 104)
Relative humidity [%]
30 ~ 85
Atmospheric pressure [kPa]
80 ~ 109
1.
Please be aware that the bending angle can be affected if the insertion cord is not kept straight.
2.
There is no guarantee that accessories selected solely using this minimum instrument channel
width will be compatible in combination.
3.
Storage under higher temperatures may impact shelf life.
6. Trouble shooting
If problems occur with the system, please use this trouble shooting guide to identify the cause
and correct the error.
Problem
Possible cause
Recommended action
No live image on
the left side of the
screen but User
Interface is present
on the display or the
image shown to the
left is frozen.
The endoscope not
connected to
the displaying unit.
Connect an endoscope to the blue port
on the displaying unit.
The displaying unit
and the endoscope
have communication
problems.
Restart the displaying unit by pressing
the power button for at least 2 seconds.
When the displaying unit is off restart by
pressing power button once more.
The endoscope
is damaged.
Replace the endoscope with a
new one.
A recorded image is
shown in the yellow file
management tab.
Return to live image by pressing the blue
live image tab or restart the displaying
unit by pressing the power button for at
least 2 seconds. When the displaying unit
is off restart by pressing power button
once more.
Low picture quality.
Blood, saliva etc. on the
lens (distal tip).
Gently rub the distal tip against the
mucosa. If the lens cannot be cleaned
this way remove the endoscope and
wipe the lens with sterile gauze.
Summary of Contents for aScope 4 Broncho Large
Page 17: ...BG 17 5 4 6 6 4 1 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Ambu Ambu 24...
Page 18: ...18 25 26 27 28 1 2 3 4 ETT 5 6 7 8 9 10 85 kPa 638 mmHg 11 1 5...
Page 21: ...BG 21 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Luer Lock B 15 16 17 18...
Page 28: ...28 6 2 2...
Page 29: ...BG 29 aScope 4 Broncho aScope 4 Broncho...
Page 69: ...EL 69 5 4 6 6 4 1 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Ambu Ambu 24...
Page 70: ...70 25 26 27 28 1 2 3 4 5 6 7 8 9 10 85 kPa 638 mmHg 11 1 5 1 6...
Page 73: ...EL 73 A aScope 4 Broncho 1 2 3 4 5 6 7 8 9 LED 10 11 12 13 14 Luer Lock B 15 flip top 16...
Page 80: ...80 6 2 2...
Page 81: ...EL 81 aScope 4 Broncho aScope 4 Broncho...
Page 180: ...180 aScope 4 Broncho aScope 4 Broncho...
Page 269: ...RU 269 5 4 6 6 4 1 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Ambu Ambu...
Page 270: ...270 24 25 26 27 28 1 2 3 4 ETT 5 6 7 8 9 10 85 638 11 1 5...
Page 273: ...RU 273 1 2 3 4 5 6 7 8 9 10 11 12 13 14 B 15 16 17 18 C...
Page 280: ...280 6 2 2...
Page 281: ...RU 281 aScope 4 Broncho aScope 4 Broncho...
Page 339: ...ZH 339 6 2 2 aScope 4 Broncho aScope 4 Broncho...