![Agilent Technologies E5070B User Manual Download Page 295](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065295.webp)
Chapter 7
289
Fixture Simulator
Determining Characteristics After Adding a Matching Circuit
7
. F
ixt
ur
e
Sim
ula
to
r
Step 5.
Press
Select Circuit
.
Step 6.
Select a matching circuit model (Figure 7-5).
Softkey
Function
None
No matching circuit is added.
Series L - Shunt C
Selects a circuit model consisting of a series inductor and a shunt
capacitor
Shunt C - Series L
Selects a circuit model consisting of a shunt capacitor and a series
inductor
Shunt L - Series C
Selects a circuit model consisting of a shunt inductor and a series
capacitor
Series C - Shunt L
Selects a circuit model consisting of a series capacitor and a shunt
inductor
Shunt L - Shunt C
Selects a circuit model consisting of a shunt inductor and a shunt
capacitor
User
Selects the circuit model defined in the user file imported by
performing Step 4
Step 7.
Specify the values of the elements in the selected circuit model.
Softkey
Function
C
Specifies capacitance [F]
G
Specifies conductance [S]
L
Specifies inductance [H]
R
Specifies resistance [
Ω
]
NOTE
When either of the two components of Series-C, "C" and "G", is set to 0, the component is
defined as "Open"; when both components are set to 0, they are defined as "Short."
When either of the two components of Shunt-L, “L” and “R”, is set to 0, the component is
defined as "Short"; when both components are set to 0, they are defined as "Open."
Step 8.
Repeat Step 3 to Step 7 to set up the matching circuit for each port used.
Step 9.
Press
Port Matching
to turn the matching circuit function
ON
.
Step 10.
Press
Return
.
Step 11.
If
Fixture Simulator
is
OFF
, press the key again to turn it
ON
.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...