108
Chapter 4
Calibration
OPEN/SHORT Response Calibration (reflection test)
Step 4.
Select OPEN or SHORT response calibration.
Softkey
Function
Response (Open)
Displays softkeys for performing an OPEN response calibration
(response calibration with an OPEN standard)
Response (Short)
Displays softkeys for performing a SHORT response calibration
(response calibration with a SHORT standard)
Step 5.
Press
Select Port
.
Step 6.
Select the test port upon which you will perform OPEN/SHORT response calibration.
Softkey
Function
1
Selects port 1
2
Selects port 2
3
Selects port 3
4
Selects port 4
NOTE
An asterisk (*) in the upper-right of the softkey indicates that the existing calibration
coefficient will be cleared if you select the test port and execute the acquisition of the
calibration coefficient (pressing
Done
).
Step 7.
Depending on the selection made in Step 4, connect an OPEN or SHORT calibration
standard to the test port (connector to which the DUT is to be connected) selected in Step 6.
Figure 4-5
Connecting the standard at OPEN/SHORT response calibration
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...