Chapter 12
409
Optimizing Measurements
Reducing Measurement Error in High Temperature Environments
1
2
. O
pt
imiz
ing
M
e
as
ur
em
e
nt
s
Reducing Measurement Error in High Temperature
Environments
The E5070B/E5071B is designed to obtain the best measurement accuracy at the ambient
temperature range of 23
°
C
±
5
°
C. The high temperature measurement mode of the
E5070B/E5071B reduces measurement error (drift error) at an ambient temperature of
28
°
C to 33
°
C.
NOTE
The high temperature measurement mode must be turned off when the analyzer is used at
an ambient temperature below 28
°
C. Otherwise, the measurement accuracy may be
degraded.
Procedure
Step 1.
Press
.
Step 2.
Press
Service Menu
.
Step 3.
Press
High Temperature
to turn
ON/OFF
the high temperature measurement mode.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...