Chapter 4
125
Calibration
Full 4-Port Calibration
4. Cal
ibra
tion
Figure 4-20
Connecting standards in full 4-port calibration
Procedure
Step 1.
Press
or
to select the channel for which you want to perform the
calibration.
Step 2.
Press
.
Step 3.
Press
Calibrate
.
Step 4.
Press
4-Port Cal
.
Step 5.
Press
Reflection
.
Step 6.
Connect an OPEN calibration standard to test port 1 (the connector to which the DUT is to
be connected).
Step 7.
Press
Port 1 Open
to start the calibration measurement.
Step 8.
Disconnect the OPEN calibration standard connected in Step 6 and replace it with a
SHORT calibration standard.
Step 9.
Press
Port 1 Short
to start the calibration measurement.
Step 10.
Disconnect the SHORT calibration standard connected in Step 8 and replace it with a
LOAD calibration standard.
Step 11.
Press
Port 1 Load
to start the calibration measurement.
Step 12.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...