228
Chapter 4
Calibration
Vector-Mixer Calibration
measured mixer. In vector-mixer calibration, where the up/down conversion method is
used, the power of the LO signal is distributed to the calibration mixer and the measured
mixer through the power splitter. During a characteristics evaluation of the calibration
mixer, the LO power level used by the drive of the calibration mixer must be equal to the
LO power level with the measured mixer connected. This is because the mixer’s
conversion loss and reflection coefficient are significantly affected by the power level of
the LO signal.
Step 1.
Select
CalKit
(3 in Figure 4-67).
Step 2.
Select the type number of the calibration kit from
CalKit
menu (3 in Figure 4-67).
Step 3.
Press the
Open
button (5 in Figure 4-67) to start measurement in OPEN.
Step 4.
Press the
Short
button (6 in Figure 4-67) to start measurement in SHORT.
Step 5.
Press the
Load
button (7 in Figure 4-67) to start measurement in LOAD.
7. Measuring Calibration Mixer with IF Filter (when using ECal module)
Figure 4-69
Characterizing calibration mixer using ECal
Step 6.
Select
ECal
(1 in Figure 4-69).
Step 7.
Select the port used for the ECal module (2 in Figure 4-69).
Step 8.
Press the
Measure
button (3 in Figure 4-69) to start measurement.
8. Saving Characteristic Data and Closing VBA Macro
Step 1.
Press the Save button (8 in Figure 4-67) to open the Save screen (Figure 4-70).
Step 2.
Press the Save button (1 in Figure 4-70) to specify a name for the characteristic data of the
calibration mixer with IF filter. Then save it to a Touchstone file.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...