560
Chapter 16
Specifications and Supplemental Information
Data accuracy enhancement
Data accuracy enhancement
Measurement calibration
Measurement calibration significantly reduces measurement uncertainty
due to errors caused by system directivity, source and load match,
tracking and crosstalk. Full 2-port, 3-port, or 4-port calibration removes
all the systematic errors for the related test ports to obtain the most
accurate measurements.
Calibration types available
Response
Simultaneous magnitude and phase correction of frequency response
errors for either reflection or transmission measurements.
Response and isolation
Compensates for frequency response and crosstalk errors of
transmission measurements.
Enhanced response
Compensates for frequency response and source match errors.
One-port calibration
Available on test port 1, port 2, port 3, or port 4 to correct for directivity,
frequency response and source match errors.
Full 2-port/3-port/4-port calibration
TRL calibration
Compensates for directivity, source match, reflection tracking, load
match, transmission tracking and crosstalk. Crosstalk calibration can be
omitted.
Interpolated error correction
With any type of accuracy enhancement applied, interpolated mode
recalculates the error coefficients when the test frequencies are changed.
The number of points can be increased or decreased and the start/stop
frequencies can be changed.
Velocity factor
Enter the velocity factor to calculate the equivalent physical length.
Reference port extension
Redefine the measurement plane from the plane where the calibration
was done.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...