![Agilent Technologies E5070B User Manual Download Page 145](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065145.webp)
Chapter 4
139
Calibration
2-port TRL calibration
4. Cal
ibra
tion
that will be cleared if you select the test port and execute acquisition of the calibration
coefficient (by pressing
Done
).
Step 7.
Thru/Line
to start the measurement of the calibration standard.
Softkey
Function
Port x-y Thru
Executes thru/line calibration for test ports x and y.
Step 8.
Reflect
to start the measurement of the calibration standard.
Softkey
Function
Port x Reflect
Executes reflection calibration for test port x.
Port y Reflect
Executes reflection calibration for test port y.
Step 9.
Line/Match
to start the measurement of the calibration standard.
Softkey
Function
x-y Line/Match
*1.This is equivalent to measuring
x-y Fwd (Syx)
and
x-y Rvs (Sxy)
.
*1
Executes line/match calibration for test ports x and y. Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
x-y Fwd (Syx)
*2.Measuring
x-y Fwd (Syx)
and
x-y Rvs (Sxy)
is equivalent to measuring
x-y Line/Match
.
*2
Executes line/match calibration for test ports x and y. Use this to perform
forward-direction measurement only.
x-y Rvs (Sxy)
Executes line/match calibration for test ports x and y. Use this to perform
reverse-direction measurement only.
Step 10.
Press
Done
to finish TRL 2-port calibration. At this point, the calibration coefficient is
calculated and saved. The error correction function is automatically turned on.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...