506
Chapter 15
Measurement Examples
Measuring the Deviation from a Linear Phase
2. Define the Measurement Conditions
The measurement conditions are defined by following the steps described below.
Setup Description
Key Operation
Presetting
-
OK
Center frequency: 1.09 GHz
Frequency span: 20 MHz
Measurement parameter: S
21
-
S21
Data format: Expand Phase
-
Expand Phase
Executing the Auto Scale
-
Auto Scale
3. Execute the Calibration
The THRU response calibration is executed.
Setup Description
Key Operation
Executing the THRU response
(A THRU standard is connected instead of a
DUT)
-
Calibrate
-
Response (Thru)
-
Thru
-
Done
4. Connect the DUT and Execute the Auto Scale
The DUT is connected again as shown in Figure 15-21 to execute the auto scale.
Setup Description
Key Operation
Executing the auto scale
-
Auto Scale
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...