![Agilent Technologies E5070B User Manual Download Page 676](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065676.webp)
670
Appendix D
Softkey Functions
Calibration Menu
(Continued)
Calibrate
2-Port TRL Cal
Thru/Line
Displays softkeys for executing THRU/LINE calibrations. Symbols x and y refer to the two test ports selected in
the
Select Ports
menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214).
x-y Thru/Line
Executes a THRU/LINE calibration on test ports x and y.
:SENS{1-16}:CORR:COLL:TRLT x,y
:SENS{1-16}:CORR:COLL:TRLT y,x
Return
Returns to the softkey display screen one level higher.
Reflect
Displays softkeys for executing reflection calibrations. Symbols x and y refer to the two test ports selected in the
Select Ports
menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214).
Port x
Reflect
Executes a OPEN/SHORT reflection calibration on test port x.
:SENS{1-16}:CORR:COLL:TRLR x
Port y
Reflect
Executes a OPEN/SHORT reflection calibration on test port y.
:SENS{1-16}:CORR:COLL:TRLR y
Return
Returns to the softkey display screen one level higher.
Line/Match
Displays softkeys for executing LINE/MATCH calibrations. Symbols x and y refer to the two test ports selected in
the
Select Ports
menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214).
x-y
Line/Match
Executes a LINE/MATCH calibration on test ports x and y.
:SENS{1-16}:CORR:COLL:TRLL x,y
:SENS{1-16}:CORR:COLL:TRLL y,x
x-y Fwd(Syx)
Executes a forward LINE/MATCH calibration on test ports x and y.
:SENS{1-16}:CORR:COLL:TRLL y,x
x-y Rvs(Sxy)
Executes a reverse LINE/MATCH calibration on test ports x and y.
:SENS{1-16}:CORR:COLL:TRLL x,y
Return
Returns to the softkey display screen one level higher.
Done
Terminates a calibration process and calculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:SAVE
Cancel
Displays softkeys for canceling a calibration.
OK
Cancels a calibration process and clears the measurement data of
standard.
:SENS{1-16}:CORR:COLL:CLE
Cancel
Returns to the softkey display screen one level higher.
Return
Returns to the softkey display screen one level higher.
3-Port TRL Cal
Displays softkeys for executing 3-port TRL calibrations.
Select Ports
Displays softkeys for selecting test ports.
1-2-3
Selects test ports 1, 2, and 3.
:SENS{1-16}:CORR:COLL:METH:TRL3 1,2,3
1-2-4
Selects test ports 1, 2, and 4.
:SENS{1-16}:CORR:COLL:METH:TRL3 1,2,4
1-3-4
Selects test ports 1, 3, and 4.
:SENS{1-16}:CORR:COLL:METH:TRL3 1,3,4
2-3-4
Selects test ports 2, 3, and 4.
:SENS{1-16}:CORR:COLL:METH:TRL3 2,3,4
Cancel
Returns to the softkey display screen one level higher.
Thru/Line
Displays softkeys for executing THRU/LINE calibrations. Symbols x, y and z refer to the three test ports selected
in the
Select Ports
menu (x, y, and z are always 1, 2, and 3, respectively, for models with Option 313 or 314).
x-y Thru/Line
Executes a THRU/LINE calibration on test ports x and y.
:SENS{1-16}:CORR:COLL:TRLT x,y
:SENS{1-16}:CORR:COLL:TRLT y,x
x-z Thru/Line
Executes a THRU/LINE calibration on test ports x and z.
:SENS{1-16}:CORR:COLL:TRLT x,z
:SENS{1-16}:CORR:COLL:TRLT z,x
y-z Thru/Line
Executes a THRU/LINE calibration on test ports y and z.
:SENS{1-16}:CORR:COLL:TRLT y,z
:SENS{1-16}:CORR:COLL:TRLT z,y
Return
Returns to the softkey display screen one level higher
Key Operation
Function
SCPI Command
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...