Effect of frequency sweep range on response resolution
Figure 9-3 shows an example when measuring the same cable while changing the sweep
span. When measured in a narrower sweep range, the overlap between two peaks is larger
than when measured in a wider sweep range. By performing measurement in a wider
sweep range, adjacent peaks can be clearly separated, which means that the response
resolution is smaller.
Figure 9-4
Effect of frequency sweep range on resolution
The sweep range affects the width of the impulse signal and the rise time of the step signal.
The width of the impulse signal and the rise time of the step signal are inversely
proportional to the sweep range. Therefore, the wider the sweep range is, the shorter these
times are.
The resolution is equal to the width defined at the point of 50% of the impulse signal or the
rise time defined at the points of 10% and 90% of the step signal. (Figure 9-5)
Figure 9-5
Definitions of impulse width and step rise time
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...