Chapter 4
Calibration
Calibration Using 4-port ECal
Step 6.
When you want to turn ON the isolation calibration, press
Isolation
(set to
ON
).
Step 7.
Select the calibration type based on the list below.
Softkey
Function
1-Port ECal
Selects 1-port calibration
2-Port ECal
Selects full 2-port calibration
3-Port ECal
*1.Options 313, 314, 413, and 414 only
*1
Selects full 3-port calibration
4-Port ECal
*2.Options 413 and 414 only
*2
Selects full 4-port calibration
Thru ECal
Selects THRU calibration
Step 8.
If you must select a port, the softkey for making this selection is displayed. Select a port
and start calibration. If you do not have to select a port, this step is skipped.
Step 9.
The E5070B/E5071B detects the test ports connected to the ECal and then measurement
starts.
NOTE
If a test port to be calibrated is not connected to the ECal module, an error occurs.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...