Chapter 4
207
Calibration
Specifying Different Standard for Each Frequency
4. Cal
ibra
tion
Example 4-4
When a part of a frequency range specified with a subclass lies outside the
measurement frequency range
For a measurement frequency range (1 GHz - 6 GHz), if you define a standard of subclass
1 (1 GHz - 5 GHz) and a standard of subclass 2 (5 GHz-8 GHz), a part of subclass 2 (6
GHz-8 GHz) lies outside the measurement frequency range, but the other part (5 GHz-6
GHz) lies within it, so calibration is possible.
Figure 4-60
When a part of a frequency range lies outside the range
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...