![Agilent Technologies E5070B User Manual Download Page 413](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065413.webp)
Chapter 12
407
Optimizing Measurements
Setting Port Extensions and Loss Values
1
2
. O
pt
imiz
ing
M
e
as
ur
em
e
nt
s
5. Measuring the OPEN/SHORT standard and executing calculation
Calculate port extensions and loss values based on the calculation results using the
OPEN/SHORT standard.
Step 1.
Press
or
to activate the channel for which you want to set auto
port extension.
Step 2.
Press
.
Step 3.
Press
Port Extensions
.
Step 4.
Press
Auto Port Extension
.
Step 5.
If you use the OPEN standard, press
Measure OPEN
, and select the port(s) for which you
want to execute measurement. Execution is restricted to ports selected in “1. Selecting a
port(s)” on page 404.
Softkey
Function
All
Measures the OPEN standards for all ports selected in “1. Selecting a
port(s)” on page 404.
Port 1
Measures the OPEN standard for port 1.
Port 2
Measures the OPEN standard for port 2.
Port 3
*1.Only with Options 313, 314, 413, and 414.
*1
Measures the OPEN standard for port 3.
Port 4
*2.Only with Options 413 and 414.
*2
Measures the OPEN standard for port 4.
Step 6.
If you use the SHORT standard, press
Measure SHORT
, and select the port(s) for which
you want to execute measurement. Execution is restricted to ports selected in “1. Selecting
a port(s)” on page 404.
Softkey
Function
All
Measures the
SHORT
standards for all ports selected in “1. Selecting a
Port 1
Measures the
SHORT
standard for port 1.
Port 2
Measures the
SHORT
standard for port 2.
Port 3
*1.Only with Options 313, 314, 413, and 414.
*1
Measures the
SHORT
standard for port 3.
Port 4
*2.Only with Options 413 and 414.
*2
Measures the
SHORT
standard for port 4.
NOTE
If a port extension value or loss value has been set, the value is updated to the calculated
result.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...