146
Chapter 4
Calibration
4-port TRL calibration
Step 7.
Reflect
to start the measurement of the calibration standard.
Softkey
Function
Port 1 Reflect
Executes reflection calibration for test port 1.
Port 2 Reflect
Executes reflection calibration for test port 2.
Port 3 Reflect
Executes reflection calibration for test port 3.
Port 4 Reflect
Executes reflection calibration for test port 4.
Step 8.
Line/Match
to start the measurement of the calibration standard.
Softkey
Function
1-2 Line/Match
*1
Executes line/match calibration for test ports 1 and 2.Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
1-2 Fwd (S21)
*2
Executes line/match calibration for test ports 1 and 2.Use this to perform
forward-direction measurement only.
1-2 Rvs (S12)
Executes line/match calibration for test ports 1 and 2.Use this to perform
reverse-direction measurement only.
1-3 Line/Match
Executes line/match calibration for test ports 1 and 3.Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
1-3 Fwd (S31)
Executes line/match calibration for test ports 1 and 3.Use this to perform
forward-direction measurement only.
1-3 Rvs (S13)
Executes line/match calibration for test ports 1 and 3.Use this to perform
reverse-direction measurement only.
1-4 Line/Match
Executes line/match calibration for test ports 1 and 4.Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
1-4 Fwd (S41)
Executes line/match calibration for test ports 1 and 4.Use this to perform
forward-direction measurement only.
1-4 Rvs (S14)
Executes line/match calibration for test ports 1 and 4.Use this to perform
reverse-direction measurement only.
2-3 Line/Match
Executes line/match calibration for test ports 2 and 3.Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
2-3 Fwd (S32)
Executes line/match calibration for test ports 2 and 3.Use this to perform
forward-direction measurement only.
2-3 Rvs (S23)
Executes line/match calibration for test ports 2 and 3.Use this to perform
reverse-direction measurement only.
2-4 Line/Match
Executes line/match calibration for test ports 2 and 4.Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...