558
Chapter 16
Specifications and Supplemental Information
Measurement capabilities
Measurement capabilities
Number of measurement channels
Up to 16 independent measurement channels. A measurement channel is
coupled to stimulus response settings including frequency, IF
bandwidth, power level, and number of points.
Number of display windows
Each measurement channel has a display window. Up to 16 display
windows (channels) can be displayed.
Number of traces
Four display modes (selectable)
16 data traces and 16 memory traces per channel at 4-channel mode
9 data traces and 9 memory traces per channel at 9-channel mode
6 data traces and 6 memory traces per channel at 12-channel mode
4 data traces and 4 memory traces per channel at 16-channel mode
Measurement choices
Opt. 213/214: S
11
, S
21
, S
12
, S
22
Opt. 313/314: S
11
, S
21
, S
31
, S
12
, S
22
, S
32
, S
13
, S
23
, S
33
, Mixed-mode
S-parameters, Balance parameters, CMRR
Opt. 413/414: S
11
, S
21
, S
31
, S
41
, S
12
, S
22
, S
32
, S
42
, S
13
, S
23
, S
33
, S
43
,
S
14
, S
24
, S
34
, S
44
, Mixed mode S-parameters, Balance parameters,
CMRR
Measurement parameter conversion
Available to convert S-parameters into reflection impedance,
transmission impedance, reflection admittance, transmission
admittance, and 1/S.
Data formats
Log magnitude, linear magnitude, phase, extended phase, positive
phase, group delay, SWR, real, imaginary, Smith chart, polar.
Data markers
10 independent markers per trace. Reference marker available for delta
marker operation. Smith chart format includes 5 marker formats: linear
magnitude/phase, log magnitude/phase, real/imaginary, R + jX, and G +
jB. Polar chart format includes 3 marker formats: linear
magnitude/phase, log magnitude/phase, and real/imaginary.
Marker functions
Marker search
Max value, Min value, peak, peak left, peak right, target, target left,
target right, bandwidth parameters with user-defined bandwidth values.
Marker-to functions
Set start, stop, center to active marker stimulus value; set reference to
active marker response value; set electrical delay to group delay at
active marker.
Search range
User definable.
Tracking
Performs marker search continuously or on demand.
Time domain functions
Transformation
Selectable transformation type from bandpass, lowpass impulse,
lowpass step.
Selectable window from maximum, normal and minimum.
Gated functions
Selectable gated filter type from bandpass, notch.
Selectable gate shape from maximum, normal and wide.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...