Chapter 5
249
Making Measurements
Setting the Averaging Trigger Function
5. M
a
king
Me
asur
eme
n
t
Setting the Averaging Trigger Function
The averaging trigger function is used to execute the sweep the number of times specified
by the averaging factor with a single trigger when the sweep averaging function is ON.
Averaging trigger
Function
On
Performs the sweep the number of times specified by the
averaging factor with a single trigger.
Off
Performs the sweep once with a single trigger.
The averaging factor is cleared before the start of measurement.
NOTE
When the point trigger function is on, its setting has priority, and you need to generate
triggers based on “(number of measurement points)
×
(averaging factor)”.
When the sweep averaging function is off, sweep is performed only once even if the
averaging trigger function is set to on.
The averaging trigger function is valid for all channels.Note that you can set the sweep
averaging function for each channel.
Using the averaging trigger function
1. Setting the averaging trigger function
When the sweep averaging function is on, follow these steps to set the averaging trigger
function.
Step 1.
Press
.
Step 2.
Press
Avg Trigger
.
Softkey
Function
ON
Sets the averaging trigger function to ON.
OFF
Sets the averaging trigger function to OFF.
2. Executing averaging measurement
Step 1.
Press
.
Step 2.
Press
Single
. The averaging factor is cleared before the start of measurement, the sweep is
executed the number of times specified by the averaging factor, and then the instrument
waits for the next trigger.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...