![Agilent Technologies E5070B User Manual Download Page 503](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065503.webp)
Chapter 15
497
Measurement Examples
Evaluating a Duplexer
15
. Me
asur
eme
n
t Exa
m
ple
s
Figure 15-11
Connecting the ECal module
Step 8.
The EcalAssistant (complete) dialog box (Figure 15-12) appears. Press the
Done
button to
terminate the calibration.
Figure 15-12
ECalAssistant (complete) dialog box
5. Connect the DUT
The DUT is connected to test ports 1, 2, and 3.
Figure 15-13
Connecting the DUT