Calibration
Calibration between Ports of Different Connector Types
4. Cal
ibra
tion
based on the setting in “4. Setting Calibration Kit” on page 157 appears (Figure 4-39).
Figure 4-39
Display of calibration kit setting for adapter calibration mode
The calibration procedure in the adapter calibration mode is the same as that in the normal
calibration except that the standard connected for each calibration data measurement
differs. Therefore, when performing this calibration, you must use the same softkeys (the
menu displayed by
-
Calibrate
) you used in the normal calibration.
In the adapter calibration mode, the standard name is displayed in the softkey to perform
each calibration data measurement based on the setting in “4. Setting Calibration Kit” on
page 157.
Following the on-screen information in Figure 4-39 and the softkey label, connect the
appropriate standard and measure each type of calibration data to perform calibration.
Step 3.
After calibration, press
Exit
to return the E5070B/E5071B to the normal calibration mode.
6. Closing VBA Macro
Press
Close
(5 in Figure 4-36) to close the macro.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...