![Agilent Technologies E5070B User Manual Download Page 162](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065162.webp)
156
Chapter 4
Calibration
Calibration between Ports of Different Connector Types
Operating procedure
NOTE
This VBA macro changes the definition of the label of calibration kit 10 (calibration kit
corresponding to the lowest softkey) temporarily, performs calibration, and restores the
previous definition after completing calibration. Therefore, if the VBA macro is aborted
for some reason, the definition of calibration kit 10’s label may be lost.
When you use this VBA macro, it is recommended that you do not use calibration kit 10. If
you are, you should back up calibration kit 10 before using the VBA macro.
To back up a calibration kit, use the VBA macro “SavRecCalKit.vba”. You can download
this VBA macro from our product information Web site established for the Agilent
Technologies E5070B/E5071B.
1. Setting Stimulus Condition
Set the stimulus condition of the channel for which you will perform the calibration.
For information on the setting procedure, see Chapter 3, “Setting Measurement
Conditions.”
2. Starting VBA Macro
Load the VBA project and run it.
Step 1.
Press
.
Step 2.
Press
Load Project
.
Step 3.
The Open dialog box appears. Specify the file name
“D:\Agilent\AdapterCharacterization.vba” and press
Open
.
Step 4.
Press
to start the macro (Figure 4-36).
Figure 4-36
Adapter Characterization macro
3. Selecting a Channel
Select a channel (1 in Figure 4-36).
NOTE
The selected channel has no relation to the active channel.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...