Calibration
Unknown Thru Calibration
Procedure to select the standard type and define the standard factor
Step 1.
Press
.
Step 2.
Press
Modify Calkit
.
Step 3.
Press
Define STDs
.
Step 4.
Press
{Defined Name}
.
Step 5.
Press
STD Type
.
Step 6.
Press
Unknown Thru
.
Step 7.
Press
Offset Delay
, then enter approximate offset delay value of the unknown thru
standard.
Procedure to define the unknown thru
Step 1.
Press
.
Step 2.
Press
Modify Calkit
.
Step 3.
Press
Specify CLSs
.
Step 4.
Press
Thru
.
Step 5.
Press
Set All
.
Step 6.
Press
{Defined Name}
.
NOTE
Different unknown thru can be used for different ports.
Procedure to perform the unknown thru
Step 1.
Press
.
Step 2.
Press
Calibrate
.
Step 3.
Press
n-Port Cal
*1
.
Step 4.
Press
Transmission
.
Step 5.
Press
{Defined Name}
to execute mechanical unknown thru calibration.
Step 6.
Complete Reflection, and Isolation (optional) for n-full port calibration.
*1.The 1-port calibration is not available.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...