Chapter 11
385
Limit Test
Ripple Test
11
. Lim
it T
es
t
Step 1.
Press
.
Step 2.
Press
Ripple Limit
.
Step 3.
Press
Fail Sign
. This menu toggles between on and off.
In addition to the screen, the following features also let you confirm the test results:
•
Beep notifying that the result was “fail”
•
Status register (for further information, see the
Programmer's Guide
)
Configuring ripple limit
You must configure the ripple limit before you can use the ripple test function. You can
specify a ripple limit table for each trace, where up to 12 ripple limit bands (frequency
bands) can be configured.
Operational procedure
Follow the steps below to configure the ripple limits.
Step 1.
Press
or
to activate the channel to which you want to apply the
ripple test function.
Step 2.
Press
or
to activate the trace to which you want to apply the ripple
test function.
Step 3.
Press
to display the Analysis menu.
Step 4.
Press
Ripple Limit
to display the softkeys for to the ripple test.
Step 5.
Press Edit Ripple Limit to display the ripple limit table shown in Figure 11-9 on page 386,
and create or edit the bands. Note that no frequency band is provided in the ripple limit
table by default. The Edit Ripple Limit menu is also displayed, which can be used for
creating or editing the frequency band. The following are functions corresponding to each
softkey.
Softkey
Function
Delete
Deletes the frequency band with the selected cell in the ripple limit table.
Add
Adds a frequency band ahead of the one for the selected cell in the ripple
limit table.
Clear Limit Table
Clears all of the data within the ripple limit table.
Export to CSV File
Outputs the ripple limit table to a file in the csv format. For further
information, see “Saving/recalling ripple limit table” on page 387.
Import from CSV File
Calls the ripple limit table saved in the csv format. For further
information, see “Saving/recalling ripple limit table” on page 387.
Return
Ends the Edit Ripple Limit menu to close the ripple limit table.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...