Chapter 4
Calibration
Enhanced Response Calibration
Figure 4-14
Connecting the Standard at Enhanced Response Calibration
Procedure
Step 1.
Press
or
to select the channel for which you want to perform the
calibration.
Step 2.
Press
.
Step 3.
Press
Calibrate
.
Step 4.
Press
Enhanced Response
.
Step 5.
Press
Ports
to select the test ports on which an enhanced response calibration will be
performed.
Softkey display
Function
2-1(S21 S11)
Measure between test ports 1 and 2. selecting port 1 as the output.
3-1(S31 S11)
Measure between test ports 1 and 3. selecting port 1 as the output.
4-1(S41 S11)
Measure between test ports 1 and 4. selecting port 1 as the output.
1-2(S12 S22)
Measure between test ports 2 and 1. selecting port 2 as the output.
3-2(S32 S22)
Measure between test ports 2 and 3. selecting port 2 as the output.
4-2(S42 S22)
Measure between test ports 2 and 4. selecting port 2 as the output.
1-3(S13 S33)
Measure between test ports 3 and 1. selecting port 3 as the output.
2-3(S23 S33)
Measure between test ports 3 and 2. selecting port 3 as the output.
4-3(S43 S33)
Measure between test ports 3 and 4. selecting port 3 as the output.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...