![Agilent Technologies E5070B User Manual Download Page 93](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065093.webp)
Chapter 3
87
Setting Measurement Conditions
Setting the Scales
3
. Set
ting
M
e
as
ur
em
e
nt
C
onditions
Manual scale adjustment on the Smith chart/polar format
Manual scale adjustment on the Smith chart format or the polar format is done by using the
displacement (
Scale/Div
of the outermost circle, Figure 3-12).
Figure 3-12
Manual scale setup on the Smith chart/polar format
Manually setting scales on the Smith chart/polar format
Step 1.
Press
(or
) and
(or
) to select the trace for
which the scale will be adjusted.
Step 2.
Press
.
Step 3.
Press
Scale/Div
.
Step 4.
Using the ENTRY block keys on the front panel, input the displacement of the outermost
circle.
Setting the value of a reference line using the marker
When using a rectangular display format, it is possible to change the reference line value to
be equal to the response value of the active marker on the active trace.
Setting the reference line value using the marker
Step 1.
Place the active marker on the active trace on the position that corresponds to the new
reference line value.
Step 2.
Press
or
.
Step 3.
Press
Marker
→
Reference
to change the reference line value to the marker response
value.
NOTE
If the reference marker is on and the stimulus value of the active marker is expressed using
a value relative to the reference marker, the absolute stimulus value will be used to set the
new reference line value.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...