126
Chapter 4
Calibration
Full 4-Port Calibration
Step 13.
Repeat Step 6 to Step 11 on test port 3.
Step 14.
Repeat Step 6 to Step 11 on test port 4.
Step 15.
Press
Return
.
Step 16.
Press
Transmission
.
Step 17.
Make a THRU connection between ports 1 and 2 (between the connectors to which the
DUT will be connected).
Step 18.
Press
Port 1-2 Thru
to start the calibration measurement.
Step 19.
Repeat Step 17 and Step 18 on ports 1 and 3.
Step 20.
Repeat Step 17 and Step 18 on ports 1 and 4.
Step 21.
Repeat Step 17 and Step 18 on ports 2 and 3.
Step 22.
Repeat Step 17 and Step 18 on ports 2 and 4.
Step 23.
Repeat Step 17 and Step 18 on ports 3 and 4.
Step 24.
Press
Return
.
Step 25.
If an isolation calibration must be performed using a LOAD standard, follow the procedure
below.
a.
Press
Isolation (Optional)
.
b.
Connect a LOAD standard to each of the four test ports (connectors to which the DUT
is to be connected).
c.
Press
Port 1-2 Isol
to start the calibration measurement.
d.
Press
Port 1-3 Isol
to start the calibration measurement.
e.
Press
Port 1-4 Isol
to start the calibration measurement.
f.
Press
Port 2-3 Isol
to start the calibration measurement.
g.
Press
Port 2-4 Isol
to start the calibration measurement.
h.
Press
Port 3-4 Isol
to start the calibration measurement.
Step 26.
Press
Return
.
Step 27.
Press
Done
to terminate the full 4-port calibration process. Upon pressing this key,
calibration coefficients will be calculated and saved. The error correction function will also
be automatically enabled. By pressing
Done
, previously saved calibration coefficients will
be overwritten with new ones.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...