Calibration Using 4-port ECal
The E5070B/E5071B allows you to perform calibration using the 4-port ECal module. It
provides much simpler operation than when using the 2-port ECal. Especially when using a
multi-port test set, calibration time and operator errors can be reduced significantly.
Operational procedure
To execute full 2-port calibration using the 4-port ECal module, follow these steps.
Step 1.
Connect the USB cable between the USB port of the 4-port ECal module and the USB port
of the E5070B/E5071B. You can make this connection while the E5070B/E5071B’s power
is on.
Step 2.
Connect the ports of the 4-port ECal module to the test ports you want to calibrate.
Figure 4-29
Connecting 4-port ECal module (for full 4-port calibration)
NOTE
You can connect the ports of the ECal and the test ports of the E5070B/E5071B
arbitrarily.Connected ports can be manually specified although they are automatically
detected before data measurement.
For more information, “Turning off ECal auto-detect
NOTE
If you don’t use all of the ECal module’s ports, connect terminations to the unused ports.
Step 3.
Press
or
to select the channel for which you want to perform the
calibration.
Step 4.
Press
.
Step 5.
Press
ECal
.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...