Chapter 7
Fixture Simulator
Converting Reference Impedance of Balanced Port
Converting Reference Impedance of Balanced Port
By using the port impedance conversion function, you can specify the impedance of each
test port. As a result of this conversion, the impedance of the balanced port in differential
mode is set to a value twice as large as the impedance of the two unbalanced ports before
conversion, and in common mode to a value one-half as large (Figure 7-17).
NOTE
Be sure to set the impedances of the two unbalanced ports equal to each other. For more
details on setting up port impedance for unbalanced ports, refer to “Converting the Port
Impedance of the Measurement Result” on page 286.
Figure 7-17
Port impedance after a balance-unbalance conversion
As described above, the impedance of the balanced port is automatically specified as the
result of specifying the impedance of the two unbalanced ports prior to balance-unbalance
conversion. However, the port impedance can be changed to an arbitrary value by using the
differential port impedance conversion function and the common port impedance
conversion function.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...