Chapter 2
51
Overview of Functions
Screen Area: Names and Functions of Parts
2.
Overvie
w
of Fun
ction
s
5-10. Channel Measurement Status
Displays the update status of traces on the channel.
!
Measurement in progress. When the sweep time exceeds
1.5 seconds,
↑
is displayed at the point on the trace.
#
Invalid traces. The measurement conditions have changed,
but the traces on the channel currently displayed have not
been updated to match the new conditions.
(No display)
The measurement has not been executed.
5-11. Error Correction Status
Displays the execution status of error correction on the channel. For details, see “Execution
status of error correction for each channel” on page 102.
5-12. Power Calibration Status
Displays the execution status of power level error correction on the channel. For details,
see “Turning ON or OFF power level error correction” on page 208.
5-13. Port Extension Status
Shows whether the port extension is turned ON or OFF.
PExt
(displayed in blue)
Port extension: ON
(not displayed)
Port extension: OFF
5-14. Fixture Simulator Status
Shows whether the fixture simulator is turned ON or OFF.
Sim
(displayed in blue)
Fixture simulator: ON
(not displayed)
Fixture simulator: OFF
5-15. Averaging Status
Displays the averaging factor and averaging count when averaging is turned on.
n/m
(displayed in blue)
Averaging: ON
(m: averaging factor; n: averaging count)
(not displayed)
Averaging: OFF
5-16. Sweep Range
Indicates the sweep range by using the start/stop or center/span.
5-17. IF Bandwidth/CW Frequency
Indicates the IF bandwidth when the sweep type is linear/log frequency or the CW
frequency when the sweep type is power.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...