666
Appendix D
Softkey Functions
Calibration Menu
(Continued)
Calibrate
1-Port Cal
Displays softkeys for executing 1-port calibrations.
Select Port
Displays softkeys for selecting a test port.
1
Selects test port 1.
:SENS{1-16}:CORR:COLL:METH:SOLT1 1
2
Selects test port 2.
:SENS{1-16}:CORR:COLL:METH:SOLT1 2
Selects test port 3.
:SENS{1-16}:CORR:COLL:METH:SOLT1 3
Selects test port 4.
:SENS{1-16}:CORR:COLL:METH:SOLT1 4
Select Port
Cancel
Returns to the softkey display screen one level higher.
Open
Executes an OPEN calibration on the test port selected in the
Select
Port
menu.
:SENS{1-16}:CORR:COLL:OPEN
Short
Executes a SHORT calibration on the test port selected in the
Select
Port
menu.
:SENS{1-16}:CORR:COLL:SHOR
Load
Executes a LOAD calibration on the test port selected in the
Select
Port
menu.
:SENS{1-16}:CORR:COLL:LOAD
Done
Terminates a calibration process and calculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:SAVE
Cancel
Displays softkeys for canceling a calibration.
OK
Cancels a calibration process and clear the measurement data of
standard.
:SENS{1-16}:CORR:COLL:CLE
Cancel
Returns to the softkey display screen one level higher.
Return
Returns to the softkey display screen one level higher.
2-Port Cal
Displays softkeys for executing a full 2-port calibration.
Select Ports
Displays softkeys for selecting test ports.
1-2
Selects test ports 1 and 2.
:SENS{1-16}:CORR:COLL:METH:SOLT2 1,2
1-3
Selects test ports 1 and 3.
:SENS{1-16}:CORR:COLL:METH:SOLT2 1,3
1-4
Selects test ports 1 and 4.
:SENS{1-16}:CORR:COLL:METH:SOLT2 1,4
2-3
Selects test ports 2 and 3.
:SENS{1-16}:CORR:COLL:METH:SOLT2 2,3
2-4
Selects test ports 2 and 4.
:SENS{1-16}:CORR:COLL:METH:SOLT2 2,4
3-4
Selects test ports 3 and 4.
:SENS{1-16}:CORR:COLL:METH:SOLT2 3,4
Cancel
Returns to the softkey display screen one level higher.
Reflection
Displays softkeys for executing reflection calibrations. Symbols x and y refer to the two test ports selected in the
Select Ports
menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214).
Port x Open
Executes an OPEN reflection calibration on test port x.
:SENS{1-16}:CORR:COLL:OPEN x
Port x Short
Executes a SHORT reflection calibration on test port x.
:SENS{1-16}:CORR:COLL:SHOR x
Port x Load
Executes a LOAD reflection calibration on test port x.
:SENS{1-16}:CORR:COLL:LOAD x
Port y Open
Executes an OPEN reflection calibration on test port y.
:SENS{1-16}:CORR:COLL:OPEN y
Port y Short
Executes a SHORT reflection calibration on test port y.
:SENS{1-16}:CORR:COLL:SHOR y
Port y Load
Executes a LOAD reflection calibration on test port y.
:SENS{1-16}:CORR:COLL:LOAD y
Return
Returns to the softkey display screen one level higher.
Key Operation
Function
SCPI Command
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...