112
Chapter 4
Calibration
THRU Response Calibration (transmission test)
Step 8.
Press
Thru
to start the calibration measurement.
Step 9.
If an isolation calibration must be performed using a LOAD standard, follow the procedure
below.
a.
Connect a LOAD standard to each of the two test ports (connectors to which the DUT
is to be connected) selected in Step 6.
Figure 4-10
Connecting the LOAD standard
b.
Press
Isolation (Optional)
to start the calibration measurement.
Step 10.
Press
Done
to terminate the response calibration (and the LOAD isolation calibration)
process. Upon pressing this key, calibration coefficients will be calculated and saved. The
error correction function will also be automatically enabled.