Chapter 3
77
Setting Measurement Conditions
Setting Stimulus Conditions
3
. Set
ting
M
e
as
ur
em
e
nt
C
onditions
Setting power range manually
By turning off the Auto Power Range set function, you can set the power level at the
frequency sweep independently for each test port manually, within the range of
−
20 dBm
to 10 dBm (for instruments with option 214, 314, or 414, the range of
−
55 dBm to 10 dBm)
and at a resolution of 0.05 dB.
Selecting range
When option 214, 314, or 414 is installed, you can select from the following power ranges.
Selectable power ranges
−
20 dBm to 10 dBm
−
25 dBm to 7 dBm
−
30 dBm to 2 dBm
−
35 dBm to
−
3 dBm
−
40 dBm to
−
8 dBm
−
45 dBm to
−
13 dBm
−
50 dBm to
−
18 dBm
−
55 dBm to
−
23 dBm
NOTE
The range selection is common to all ports: You cannot select it for each port
independently.
The procedure for selecting the range is as follows:
Step 1.
Press
or
to select the channel for which you want to set the
power range.
Step 2.
Press
.
Step 3.
Press
Power
.
Step 4.
Turn OFF
Auto Range
.
Step 5.
Press
Power Ranges
.
Step 6.
Press the desired softkey to select the power range.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...