Chapter 4
237
Calibration
Scalar-Mixer Calibration
4. Cal
ibra
tion
Step 9.
Connect the OPEN calibration standard to the test port x (connector for the DUT) you have
selected in Step 6.
Step 10.
Press
Port x @Freq x Open
to start measurement of the calibration standard.
Step 11.
Press
Port x @Freq y Open
to start measurement of the calibration standard.
Step 12.
Disconnect the OPEN calibration standard you connected in Step 9 and then connect
SHORT calibration standard in its place.
Step 13.
Press
Port x@Freq x Short
to start measurement of the calibration standard.
Step 14.
Press
Port x@Freq y Short
to start measurement of the calibration standard.
Step 15.
Disconnect the SHORT calibration standard you connected in Step 12 and then connect
LOAD calibration standard in its place.
Step 16.
Press
Port x@Freq x Load
to start measurement of the calibration standard.
Step 17.
Press
Port x@Freq y Load
to start measurement of the calibration standard.
Step 18.
Follow steps Step 9 to Step 17 for port y.
Step 19.
Press
Return
.
Step 20.
Press
Transmission
.
Step 21.
Connect a THRU between the test ports x and y (between the connectors for the DUT) you
selected in Step 6.
Step 22.
Press
Port y-x@Freq y Thru
to start measurement of the calibration standard.
Step 23.
Press
Port x-y@Freq x Thru
to start measurement of the calibration standard.
Step 24.
Press
Return
.
Step 25.
Press
Power Meter
.
NOTE
In scalar-mixer calibration, the power meter is used. For more on how to set the power
meter, see “Preparing to control the power meter” on page 210.
Step 26.
Press
Use Sensor
. Each time the key is pressed, Channels A and B switch over alternately.
If you are using a one-channel power meter, select Channel A.
Step 27.
Connect the power sensor for the selected channel to the selected port, as shown in Figure
4-77.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...