206
Chapter 4
Calibration
Specifying Different Standard for Each Frequency
Example 4-2
When frequency ranges specified with subclasses do not cover a part of a
measurement frequency range
For a measurement frequency range (1 GHz - 8 GHz), if you execute calibration with a
standard of subclass 1 (1 GHz-5 GHz) and then with a standard of subclass 2 (6 GHz-8
GHz),
Done
is not available for the undefined portion (5 GHz-6 GHz).
Figure 4-58
When a part of a frequency range is not covered
Example 4-3
When a frequency range specified with a subclass lies outside the measurement
frequency range
For a measurement frequency range (1 GHz - 5 GHz), if you define a standard of subclass
1 (1 GHz - 5 GHz) and a standard of subclass 2 (5 GHz-8 GHz), calibration possible for
subclass 2 although it is out of the measurement frequency range. The calibration
coefficients, however, are disregarded. (There is no check mark on the softkey.) Note that,
because the standard of subclass 1 covers the measurement frequency range,
Done
is
available.
Figure 4-59
When a subclass is out of a frequency range
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...