Chapter 4
147
Calibration
4-port TRL calibration
4. Cal
ibra
tion
Step 9.
Press
Done
to finish TRL 4-port calibration. At this point, the calibration coefficient is
calculated and saved. The error correction function is automatically turned on.
2-4 Fwd (S42)
Executes line/match calibration for test ports 2 and 4.Use this to perform
forward-direction measurement only.
2-4 Rvs (S24)
Executes line/match calibration for test ports 2 and 4.Use this to perform
reverse-direction measurement only.
3-4 Line/Match
Executes line/match calibration for test ports 3 and 4. Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
3-4 Fwd (S43)
Executes line/match calibration for test ports 3 and 4. Use this to perform
forward-direction measurement only.
3-4 Rvs (S34)
Executes line/match calibration for test ports 3 and 4. Use this to perform
reverse-direction measurement only.
*1.This is equivalent to measuring
1-2 Fwd (S21)
and
1-2 Rvs (S12)
.
*2.Measuring
1-2 Fwd (S21)
and
1-2 Rvs (S12)
is equivalent to measuring
1-2 Line/Match
.
Softkey
Function
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...