180
Chapter 4
Calibration
Confidence Check on Calibration Coefficients Using ECal
Figure 4-54
Connecting ECal module (for verification of S21)
Step 6.
Press
.
Step 7.
Press
ECal
.
Step 8.
When using an adapter to the ECal, press
Characterization
and then press the softkey
corresponding to the characterization for the adapter you are using.
Step 9.
Press
Confidence Check
.
Step 10.
Compare the data trace and the memory trace and verify whether measurement is correct.
The following is the procedure for comparison when simultaneously displaying the data
trace and the memory trace.
a.
Press
.
b.
Press
Display
.
c.
Press
Data & Mem
.
d.
Press
.
e.
Press
Auto Scale
.
f.
Determine whether the difference between the traces is acceptable.
Step 11.
For all of the parameters you want to verify, repeat Step 3 through Step 10.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...