![Agilent Technologies E5070B User Manual Download Page 163](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065163.webp)
Chapter 4
157
Calibration
Calibration between Ports of Different Connector Types
4. Cal
ibra
tion
4. Setting Calibration Kit
Select a calibration kit for each test port and each pair of test ports and select a standard to
use for each reflection/transmission measurement.
Step 1.
Press
Specify Cal Kit
The Specify Cal Kit dialog box (Reflection tab) shown in Figure 4-37 appears.
Figure 4-37
Specify Cal Kit dialog box (Reflection tab)
Step 2.
For each test port, select a calibration kit (1 in Figure 4-37) and a standard to use in the
open/short/load reflection measurements from the calibration kits (2 in Figure 4-37).
Step 3.
Select the
Thru
tab (3 in Figure 4-37). The Specify Cal Kit dialog box (Thru tab) shown in
Figure 4-38
Specify Cal Kit dialog box (Thru tab)
Step 4.
For each test port, select a calibration kit (1 in Figure 4-38). In addition to the 10
calibration kits you can select for normal calibration, you can also select Adapter File.
When you want to use the adapter as the standard, select Adapter File.
From the calibration kits you have selected, select a standard to use for transmission
measurement (2 in Figure 4-38). If you select Adapter File as the calibration kit, you can
select an adapter file (2-port Touchstone file) under the
D:\Agilent\Data\AdapterCharacterization folder as the standard. In this case, you have to
specify the port connection mode (FWD/RVS) between the E5070B/E5071B and the
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...