![Agilent Technologies E5070B User Manual Download Page 384](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065384.webp)
378
Chapter 11
Limit Test
Limit Test
Figure 11-4
Example of limit lines (when limit table is set as Figure 11-3)
NOTE
You can define a limit line that is able to freely overlap the stimulus range of another limit
line.
Defining one limit line having the same type as a second limit line whose stimulus range
overlaps with the first one results in two or more limit values at the same measurement
point. In this case, the limit value to be used in the limit test is defined as follows:
•
When two or more limit values whose type is set to maximum (MAX) exist, the
smallest one is used as the maximum.
•
When two or more limit values whose type is set to minimum (MIN) exist, the largest
one is used as the minimum.
NOTE
Even if the span of the sweep range on the E5070B/E5071B is set to 0, enter the two
parameters of Begin Stimulus and End Stimulus.
NOTE
When two or more response values are returned as a result of using the Smith or polar chart
format, the first response value of the marker provides the object of the limit test.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...