232
Chapter 4
Calibration
Vector-Mixer Calibration
the macro (1 in Figure 4-73).
NOTE
If failure occurs when reading the data file for the calibration mixer with IF filter, the
characterization may have done by using only one port instead of using two ports.
Figure 4-73
Balanced Mixer Characterization Macro
Step 9.
Select the measurement port as shown in Figure 4-73 (2 in Figure 4-73) and then connect a
THRU between the IF ports of the calibration mixers to correct the phase difference of the
LO signals for the calibration mixers with IF filters.
Step 10.
Pressing the
Execute
(3 in Figure 4-73) button executes a phase error correction and
overwrites the results on the original data file.
NOTE
You cannot run the
Execute
function when selecting the measurement port if the data file
of the calibration mixer with IF filter (*.s2p) has not been set for the network removal
function of the fixture simulator.
NOTE
The phase error correction data reflects the phase difference of the LO signals for the phase
information of the calibration mixer’s data file, which is registered in any two ports.
Step 11.
Press the
Close
button (4 in Figure 4-73) to exit macro.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...