Chapter 4
119
Calibration
Full 2-Port Calibration
4. Cal
ibra
tion
Figure 4-16
Connecting standards in full 2-port calibration
Procedure
Step 1.
Press
or
to select the channel for which you want to perform the
calibration.
Step 2.
Press
.
Step 3.
Press
Calibrate
.
Step 4.
Press
2-Port Cal
.
Step 5.
Press
Select Ports
.
Step 6.
Select the test ports on which you will perform full 2-port calibration. (In the procedure
below, the selected test ports are denoted as x and y.)
Softkey
Function
1-2
Selects test ports 1 and 2
1-3
Selects test ports 1 and 3
1-4
Selects test ports 1 and 4
2-3
Selects test ports 2 and 3
2-4
Selects test ports 2 and 4
3-4
Selects test ports 3 and 4
NOTE
An asterisk (*) in the upper-right of the softkey indicates that the existing calibration
coefficient will be cleared if you select the test port and execute the acquisition of the
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...