![Agilent Technologies E5070B User Manual Download Page 493](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065493.webp)
Chapter 15
487
Measurement Examples
Measuring the SAW Bandpass Filter Using the Segment Sweep
15
. Me
asur
eme
n
t Exa
m
ple
s
1. Determine the Segment Sweep Conditions
Figure 15-1 shows the results of evaluating the transmission characteristics of the SAW
bandpass filter in the range of 440 MHz to 3 GHz by using the linear sweep.
Figure 15-1
Transmission characteristics of SAW bandpass filter (440 MHz to 3 GHz, linear
sweep)
The measurement conditions are determined for each frequency range. Here, the segment
sweep is performed following the sweep conditions shown in Table 15-2.
Table 15-2
Determining Sweep Conditions (using markers shown in Figure 15-1)
Frequency Range
Measurement Conditions
Start
Stop
Number of Points
IF Bandwidth
440 MHz
915 MHz
(Marker 1)
47
70 kHz
915 MHz
(Marker 1)
980 MHz
(Marker 2)
130
100 kHz
980 MHz
(Marker 2)
1.035 GHz
(Marker 3)
55
70 kHz
1.07 GHz
(Marker 4)
2 GHz
(Marker 5)
93
70 kHz
2.6 GHz
(Marker 6)
3 GHz
41
70 kHz
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...