![Agilent Technologies E5070B User Manual Download Page 353](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065353.webp)
Chapter 9
347
Analysis in Time Domain (Option 010)
Deleting Unnecessary Data in Time Domain (gating)
9.
Analy
sis in T
ime Domain
(O
ption
010
)
Operational procedure
Step 1.
Press
(or
) and
(or
) to activate a trace for
which you want to set the gate shape.
Step 2.
- Press
Gating
to display the “Gating” menu.
Step 3.
Press
Shape
and then select the gate shape from the following.
Softkey
Function
Maximum
Sets the gate shape to “maximum.”
Normal
Sets the gate shape to “normal.”
Wide
Sets the gate shape to “wide.”
Minimum
Sets the gate shape to “minimum.”
Setting gate range
Specify the gate range in time. The ends of the range are defined as the
−
6 dB attenuation
points shown in Figure 9-9. You can set the gate range by specifying the start and stop
times or the center and span. The
E5070B/E5071B
has the following limitations on the
gate range you can set.
Lower limit
−
T
span
*1
Upper limit
T
span
Operational procedure
Step 1.
Press
(or
) and
(or
) to activate the trace for
which you want to set the gate range.
Step 2.
- Press
Gating
to display the “Gating” menu.
Step 3.
Press each of the following softkeys to specify the gate range. At the side of the set value
area in the data entry bar, the distance corresponding to the set time is displayed as shown
in Figure 9-7. The displayed distance is a value that takes the velocity factor into
consideration.
Softkey
Function
Start
Sets the start time
Stop
Sets the stop time
Center
Sets the center of the gate in time
Span
Sets the gate span in time
NOTE
No Hardkey is provided for this setting. The hardkeys are dedicated to setting the sweep
*1.
T
span
is the measurement range expressed in time obtained in “Measurement range” on
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...