![Agilent Technologies E5070B User Manual Download Page 685](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065685.webp)
Appendix D
679
Softkey Functions
Calibration Menu
D. Sof
tke
y
Fu
nctions
(Continued)
Modify Cal Kit
Specify CLSs
Open
Displays softkeys for selecting standards for the OPEN standard class.
Set All
Displays softkeys for selecting standards for the OPEN standard class that applies to
None
Does not include
1. (Std Name)
in the OPEN standard class.
:SENS{1-16}:CORR:COLL:CKIT:ORD:THRU
x,y,0
1. (Std
Name)
Includes
1. (Std Name)
in the OPEN standard class.
:SENS{1-16}:CORR:COLL:CKIT:ORD:OPEN
x,1
2. (Std
Name)
Includes
2. (Std Name)
in the OPEN standard class.
:SENS{1-16}:CORR:COLL:CKIT:ORD:OPEN
x,2
.
.
.
.
.
.
21. (Std
Name)
Includes
21. (Std Name)
in the OPEN standard class.
:SENS{1-16}:CORR:COLL:CKIT:ORD:OPEN
x,21
Cancel
Returns to the softkey display screen one level higher.
Port 1
Displays softkeys for selecting standards for the OPEN standard class that applies only to test port 1. The
lower-level softkeys are the same as those for
Set All
.
Port 2
Displays softkeys for selecting standards for the OPEN standard class that applies only to test port 2. The
lower-level softkeys are the same as those for
Set All
.
Port 3
Displays softkeys for selecting standards for the OPEN standard class that applies only to test port 3. The
lower-level softkeys are the same as those for
Set All
.
Port 4
Displays softkeys for selecting standards for the OPEN standard class that applies only to test port 4. The
lower-level softkeys are the same as those for
Set All
.
Return
Returns to the softkey display screen one level higher.
Short
Displays softkeys for selecting standards for the SHORT standard
class. The lower-level softkeys displayed are the same as those for
Open
.
:SENS{1-16}:CORR:COLL:CKIT:ORD:SHOR
Load
Displays softkeys for selecting standards for the LOAD standard
class. The lower-level softkeys displayed are the same as those for
Open
.
:SENS{1-16}:CORR:COLL:CKIT:ORD:LOAD
Key Operation
Function
SCPI Command
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...